VIII. Specifications
1. Oxygen Saturation Range
(SpO
2
)
0 to 100%
2. Pulse Rate Range
18 to 300 Pulses Per Minute
3. Displays
Patient Indicator
Perfusion LED
Digital Displays
3-digit 7-segment LEDs
4. Measurement Wavelengths
Red
660 nanometers
Infrared
910 nanometers
5. Accuracy
SpO
2
(± 1 Standard Deviation)
70 to 100% ± 2 digits for
adults using Finger Clip
Sensors
70 to 100% ± 3 digits for
adults using Flex, Flexi-Form,
or Reflectance Sensors
70 to 100% ± 4 digits using
Ear Clip Sensors
70 to 95% ± 3 digits for
neonates using infant or
neonatal sensors
Below 70% is not specified for
all sensors
Pulse Rate
± 3% ± 1 digit
Standard Deviation is a statistical measure: up to 32% of the readings
may fall outside these limits.