Communications
Serial Output
The Model 920M PLUS provides real-time data output capability via the
pulse oximeter sensor connector (a 9-pin Sub-D connector). The pulse
oximeter sensor connector pin assignments are listed below.
Information from the device, in the real-time mode, is sent in an ASCII
serial format at 9600 baud with 9 data bits, 1 start bit, and 1 stop bit. The
data are output at a rate of once per second.
NOTE: The 9th data bit is used for odd parity in memory download. In
real-time mode, it is always set to the mark condition. Therefore,
real-time data may be read as 8 data bits, no parity.
Real-time data may be printed or displayed by devices other than the
pulse oximeter. On power up a header is sent identifying the format and
the date and time. Thereafter, the data are sent once per second in the
following format:
SPO2=XXX HR=YYY
where “XXX” represents the SpO
2
value, and “YYY” represents the pulse
rate. The SpO
2
and pulse rate will be displayed as “---” if there are no data
available for the data reading.
Pulse Oximeter Sensor Connector Pin Assignments
Pin
Number Assignment
1
1-Wire
®
2 Infrared Anode, Red Cathode
3 Infrared Cathode, Red Anode
4 Serial Data, TTL Levels
5 Detector Anode
6 Sensor Type
7 Cable Shield (Ground)
8 No Connection
9 Detector Cathode, +5 V